The Accordion M1 Current Source & Supply is a ready-to-run configuration of the compact Accordion M1 platform for driving and monitoring current-controlled loads, such as LEDs, in PCBA test, NPI and production test. It combines eight independently controlled constant-current outputs with per-channel current readback, two precision measurement inputs and a two-channel programmable power supply on one network-attached node.

The unit contains an Accordion M1 mainboard with a Raspberry Pi 4, pre-fitted with three modules:

  • ESH10000584 8-channel Current Source (N-Top) — eight constant-current output stages, 0–450 mA each.

  • ESH10000590 Precision ADC/DAC Module (SO-DIMM) — 16-bit current set-points and 24-bit current readback for the eight channels, plus the precision measurement inputs.

  • ESH10000662 Mini-PSU (COMTOP) — two independent programmable outputs, 5–24 V, up to 1.5 A each.

The M.2 slot is free for expansion.

The unit ships with lifetime licences for Accordion Shell and Accordion Pilot.

Typical use cases:

  • Driving and characterising LEDs and other current-controlled loads.

  • Production fixtures that need precise, software-set currents with readback.

  • Precision voltage or shunt-current measurement alongside the current outputs.

  • Supplying the DUT or a light table from the same node.

Key constraints:

  • 0–450 mA per channel; 2.4 A maximum continuous total across all eight channels.

  • Output voltage across each channel is 0–5 V and depends on the load and the output current.

  • Mini-PSU: 1.5 A continuous per channel up to 20 V output; above 20 V each channel is limited to about 30 W.

  • Powered over USB-C (USB-PD) at 20 V. The bundled power supply is the only supported supply.


Datasheet

System

Parameter

Value

Notes

Product name

Accordion M1 Current Source & Supply


Part number

ESH10000754


Platform

Accordion M1 (mainboard ESH10000597)


Controller

Raspberry Pi 4, 4 GB


Firmware

Accordion 6.0.0 or later


Power input

USB-C (USB-PD) 20 V, via the bundled 130 W power supply

Use the bundled supply only.

Power-on

At power-up or with the power button

Optional, set at delivery.

Input protection

eFuse, 5 A current limit, overvoltage cut-off above 21 V


Network interface

Ethernet, Wifi


Module slots populated

SO-DIMM (Precision ADC/DAC), N-Top (Current Source), COMTOP (Mini-PSU)


Module slots free

1× M.2


Cooling

Active, one fan, speed-regulated by the Accordion process

Full speed at boot.

Enclosure

Hammond 1455K1201BU with E# front, top and back panels


Dimensions (L × W × H)

120 × 78 × 43 mm


Weight

324 g


Compliance

CE, RoHS, REACH


Included hardware

Item

Qty

Notes

Accordion M1 unit

1

Pre-configured with the modules below.

ESH10000584 — 8-channel Current Source

1

N-Top position.

ESH10000590 — Precision ADC/DAC Module

1

SO-DIMM slot.

ESH10000662 — Mini-PSU

1

COMTOP position.

microSD card, 32 GB

1

Pre-loaded with the Accordion image.

USB-C power supply

1

130 W, USB-PD.

USB-C power cable

1

1 m.

Included software licences (lifetime)

Product

Licence

Notes

Accordion Shell

Lifetime, single-unit

CLI tool. Installed via dotnet tool install -g AccordionQ2.Shell.

Accordion Pilot

Lifetime, single-unit

Windows GUI. Installer: https://esharp.blob.core.windows.net/accordionpilot/setup.exe.

The Web API, the Python package accordionq2 and the .NET client AccordionQ2.WebApiClient are open clients and need no separate licence.

Current outputs (ESH10000584 with ESH10000590)

Parameter

Value

Notes

Channels

8, independently controlled


Output current range

0–450 mA per channel


Total output current

2.4 A maximum, continuous, all channels together


Output voltage

0–5 V

Depends on load and output current.

Set-point resolution

16-bit


Current readback

Per channel, 24-bit ADC


Current sense

0.5 Ω shunt per channel


Off-state leakage

< 1 µA


Calibration

Two-point (offset + gain) per channel, stored and applied at start-up


Precision measurement inputs (ESH10000590)

Parameter

Value

Notes

Inputs

2, single-ended or differential


Input range, single-ended

0–5 V

Scaled ÷2 to the 2.5 V ADC full scale.

Resolution

24-bit


Sensor supply

5 V, fused at 500 mA

On the measurement connector.

Power supply — Mini-PSU (ESH10000662)

Parameter

Value

Channels

2, independent

Output voltage

5–24 V per channel

Output current

1.5 A continuous per channel up to 20 V output; about 30 W per channel above 20 V

Current-limit accuracy

±5 mA after two-point calibration (0–1.5 A)

Current-limit resolution

≤ 0.75 mA

Load transient (0 → 1.5 A)

≤ 1 % overshoot, recovery ≤ 5 µs

Output stability

≤ ±0.1 % over 8 h at 24 V / 1 A

Readback

Output voltage and current per channel

Per-channel enable

Yes, software-controlled

For per-module detail see ESH10000590 and ESH10000662.


Manual

Hardware

Power-on procedure
  1. Connect the bundled USB-C power supply to mains.

  2. Connect the USB-C cable to the unit's power input.

  3. Connect Ethernet to the same network as the controlling PC.

  4. Depending on how the unit is set up, it starts when power is applied, or when you press the power button.

  5. Wait about 1 minute for the unit to boot. The fan runs at full speed until the Accordion process takes over.

  6. The unit is reachable by hostname or by its DHCP-assigned IP address.

All current outputs and the Mini-PSU outputs are off at boot and must be set and enabled in software.

Connecting loads
  • Current outputs: one 2-pin push-in connector per channel (OUT+, GND). Wire the load between OUT+ and GND.

  • Measurement inputs: the 4-pin measurement connector (5 V, GND, MEAS+, MEAS−).

  • Programmable supplies: the Mini-PSU output connector (VOUT1, VOUT2, GND).

Keep the sum of all channel currents at or below 2.4 A. Power down the unit before changing wiring.

Software

The unit runs the standard Accordion software. Use the Web API directly, or through Accordion Pilot, Accordion Shell, the Python package accordionq2, the .NET client AccordionQ2.WebApiClient or TestStand — see the Accordion A2 Lite page for install commands; the same clients apply.

  • Set each channel's current, enable it, and read back the measured current.

  • Calibration is two-point per channel; run it from Accordion Pilot or through the API. See Calibration.

  • Mini-PSU: set voltage and current limit per channel, then enable the output.


Pinout

Current outputs (8 × 2-pin, Phoenix Contact 1843606)

Pin

Signal

Description

1

OUT+

Current output, positive terminal

2

GND

Common ground

Measurement connector (4-pin, Phoenix Contact 1881574)

Pin

Signal

Description

1

5V

5 V supply, fused at 500 mA

2

GND

Ground

3

MEAS+

Single-ended input / differential positive

4

MEAS−

Single-ended input / differential negative

Maximum voltage on the measurement pins is 5 V.

Mini-PSU output connector (ESH10000662, 4-pin)

Pin

Signal

1

VOUT1

2

GND

3

GND

4

VOUT2